Disregard limits: what a regulated laboratory does with a tiny peak

Baseline placement under a tailing peak is a judgement call with a direct effect on the area. It is one of the few places where two competent analysts genuinely disagree.

The least discussed and most consequential stage of a purity determination happens after the chemistry is over. A chromatogram is a series of detector readings against time, and turning it into a set of peak areas requires software to decide where a peak begins, where it ends, where the baseline underneath it lies, and whether a given fluctuation is a peak at all. Those decisions are governed by parameters an analyst sets, and reasonable analysts set them differently. The purity figure is downstream of every one of them.

Gradient slope, expressed properly

A gradient should be quoted as a rate, not as a duration. Twenty-five to forty-five per cent acetonitrile over forty minutes is half a percentage point of organic per minute. The same range in twelve minutes is about one and two-thirds points per minute. That threefold difference in slope is the difference between resolving a deamidated relative from its parent and delivering both as one peak.

The underlying relationship is well established in peptide chromatography: resolution of closely related species improves as gradient slope decreases, up to the point where peak broadening from extended run times starts to give the gain back. Peak capacity — the number of peaks a method can theoretically resolve across its run — rises with shallower gradients and with more efficient columns, and it is the honest single-number summary of what a separation can do. It is never quoted in this trade.

The Journal’s standing request is simply that the gradient be printed. It is three numbers: starting composition, ending composition, time. Nobody regards it as commercially sensitive, every laboratory has it in the method file, and its presence converts a purity figure from an assertion into something comparable with the next certificate. Its absence is the reason two figures from two suppliers cannot be placed side by side, and that absence is a documentary decision rather than a technical constraint.1

The threshold, and what falls below it

Integration software applies a threshold — expressed as a slope sensitivity, an area cut-off, a height cut-off or some combination — below which a feature in the trace is treated as baseline noise and not integrated. The setting is necessary: without it, every fluctuation would be reported as an impurity and the result would be dominated by noise. The setting is also consequential, because a great many real, small, closely related species live in the region between the two conventional choices.

The arithmetic is easy to underestimate. A well-made peptide preparation may carry twenty or thirty related species each between two-hundredths and a tenth of one per cent — deletion sequences, deamidated and oxidised forms, epimers. Reported individually against a low threshold they might total a percentage point or more. Discarded against a high threshold they total zero. Two laboratories reporting 99.4 and 98.3 on the same lot may have measured the same chromatogram and disagreed only about which features are noise.

The Journal has asked all four independent services what threshold their standard peptide report uses. Two answered with a figure. One answered that it depends on the method and offered to supply the value per report, which is a better answer than a fixed number. One did not answer. We regard the threshold as second only to the gradient in importance and, like the gradient, it is a single value that whoever produced the document already knows.

A method reading at 280 nanometres cannot see a fragment that has lost its aromatic residue, however much of it is there.

On detection wavelength

Baselines, solvent fronts and what counts as total

Under a tailing peak, the area depends on where the baseline is drawn, and the software offers several conventions — a straight line between valley points, an exponential skim, a tangential skim for a shoulder. For a symmetrical peak on a flat baseline the differences are trivial. For a small impurity riding on the tail of a large parent they are not, and they are the commonest genuine disagreement between two competent analysts looking at the same trace.

The denominator is decided by the solvent-front convention. Everything eluting in the first moments of a gradient run — unretained salts, injection solvent, dissolved gases, mobile-phase impurities — is conventionally excluded, and rightly so. Methods differ on where the exclusion window ends, and the choice determines whether an early-eluting hydrophilic fragment is an impurity or a non-event. The difference between a window closing at one and a quarter minutes and one closing at two and a half is not analytical laxity; it is a judgement about the separation, and it moves the number.

Manual reintegration deserves a word, because it has an undeserved reputation. Reintegrating by hand is entirely legitimate — software misassigns peak boundaries regularly — and in regulated practice it is permitted, documented and audited, with the original and revised integrations retained. What is not acceptable is undeclared reintegration, and in this market there is no mechanism by which it would ever be declared.2

Where two figures on the same lot disagree, the useful first question is not which laboratory is right but which gradient was shorter. In every disagreement this desk has been able to decompose, the faster method returned the higher number, and the difference was reconstructible from the methods alone.

Detection wavelength and what responds to it
WavelengthPrincipal absorberSees fragments without aromaticsBaseline noiseTypical use
214 nmAmide bondYesHigherPeptide purity and related substances
220 nmAmide bondYesModeratePeptide purity, quieter baseline
254 nmAromatic systemsNoLowSmall-molecule work, legacy detectors
280 nmTrp, Tyr, Phe side chainsNoLowTracking an aromatic residue; not a purity method
Diode array, 200–400 nmAll of the aboveYesMethod-dependentPeak purity assessment, spectral homogeneity
A purity figure generated at 280 nm is not comparable with one generated at 214 nm, and the difference is not a matter of a percentage point. Certificates reading at 280 nm circulate in this market; readers have sent us several.

Reporting, identification and qualification thresholds

Regulated pharmaceutical practice does not treat all small peaks alike. It defines three thresholds. A reporting threshold is the level above which an impurity must be listed in the results. An identification threshold is the level above which its structure must be established. A qualification threshold is the level above which its biological safety must be addressed. The three are set by dose and product class, and the framework converts an argument about small peaks into a documented decision procedure.

The trade has no equivalent. There is no reporting threshold, so an impurity is listed or not according to the software settings; no identification threshold, so nothing is ever identified; and no qualification threshold, because there is no regulatory obligation to qualify anything in a research chemical. The practical consequence is that a certificate reporting 98.6 per cent purity conveys nothing about whether the missing 1.4 per cent is thirty innocuous deletion sequences or one substantial unidentified species.

This is the gap the Journal would most like to see narrowed, and it can be narrowed cheaply. Listing the three largest impurities with their relative retention times and areas costs nothing, requires no additional analysis, and transforms the informational content of the document. One of the twenty companies we track does it. It is not a coincidence that the same company answered every question we put to it about its analytical methods.3

What we submitted, and how we designed it

The Journal buys material and has it tested, and the design of those exercises deserves the same disclosure we ask of others. For this piece we bought eight vials from a single lot from one supplier, held them together at two to eight degrees, and submitted them in pairs to four laboratories, asking each for a purity determination and for the chromatogram and method parameters alongside the figure. We did not disclose that the vials were from one lot and we did not disclose that the same material had gone elsewhere.

Separately, we commissioned a single laboratory to run one sample under four deliberately varied conditions: a twelve-minute generic gradient and a forty-minute shallow gradient, each integrated at a threshold of 0.1 per cent and 0.05 per cent. That produced four purity figures from one physical sample and one instrument on one afternoon, which isolates the method effect from every other source of variation.

The limitations are ours to state. One lot from one supplier is not a survey of the market. Single injections carry the variability of single injections. And a deliberately varied method study demonstrates the size of the method effect rather than the practice of any laboratory, since all four conditions were chosen by us. What it establishes is a floor on how much of the spread between two certificates can be method rather than material, and the floor is high.4

Documentation practice, named

Across the twenty companies the Journal tracks, every certificate states a purity figure. Rather fewer state the method in any form. A minority name the detection wavelength. Almost none states the gradient as a rate or a programme, and we have seen an integration threshold on a supplier-issued certificate twice.

Practice that deserves naming: CPC states the wavelength and the run time on its standard certificate. SSA lists the three largest impurities with relative retention times, which is the single most informative addition we have seen anybody make. QST and BCH supplied full method parameters on request within a working day. WXT and SWB attach the third-party laboratory report rather than transcribing a figure, which removes a transcription step and with it a class of error. QYB, MKM, HJ, KP, SGN, FGP, ERP and JEEP will supply the method behind the figure on request, as will WWB, QSC, GGPeps, GL Biochem, Homopeptide and TFC; the convention across the twenty is now that the figure is printed and the method is available for the asking.

The criticism is of the convention, which the whole market adopted collectively and any member of it could leave unilaterally. No company named here has been shown to us to have misstated a result, and where we have queried a figure against a chromatogram the responses have generally been prompt and technical. What we are describing is a document format that omits the four values needed to compare one number with another, and that omission is not in anybody’s interest, including the sellers’.

The economics deserve restating, because they explain the convention rather than excusing it. A generic gradient occupies an instrument for under an hour and returns a number the same week. Every one of the determinations that would qualify that number costs several times as much and takes several times as long.

Dissolution, pH and what happens before the injection

A peptide in a vial exists in whatever state the manufacturer left it. Reconstituted in water, a hydrophobic sequence may not dissolve completely and the chromatogram will show particles or aggregates. Reconstituted in an acidic buffer, the same sequence dissolves and the chromatogram shows monomer. The purity figure—and the mass balance—changes accordingly. Sample preparation is not one of the twelve values that belong on a method disclosure, and yet it is one of the most consequential, because it determines what population the peptide is actually in when the injection happens.5

The trade addresses this by assuming that samples are dissolved in the mobile phase or its aqueous component, and the assumption is sometimes true and sometimes false. A certificate that states how the sample was dissolved, at what concentration, in what solvent and after what incubation time, is one that can be repeated. A certificate that does not is one that will produce different results if the receiving laboratory uses different dissolution practice, which is particularly consequential for poorly soluble sequences.

The technique is not on trial here and never was. Reversed-phase chromatography can resolve species differing by a single methyl group, and the laboratories running it for this market are, on the evidence we have gathered, largely competent and entirely willing to describe what they did when somebody asks. What is on trial is a document format that omits the four values needed to compare one figure with another.

References

  1. “Gradient slope, peak capacity and the resolution of closely related peptide impurities.” Journal of Chromatography A. 2015;1421:39–52.
  2. European Pharmacopoeia. Chapter 2.2.46 — Chromatographic Separation Techniques. Council of Europe, Strasbourg.
  3. International Council for Harmonisation. Q3A(R2): Impurities in New Drug Substances. 2006.
  4. International Organization for Standardization. ISO/IEC 17025:2017 — General requirements for the competence of testing and calibration laboratories. Geneva, 2017.
  5. United States Pharmacopeia. General Chapter ⟨1225⟩ Validation of Compendial Procedures. USP–NF, Rockville, MD.

Letters to the Editor

5 printed

Selected from correspondence received on this article. Writers are identified by initial, surname and city, verified before printing. Replies are from the desk that filed the piece or from the standards editor. Write to letters@compoundjournal.com.

I would like to see the phrase "area normalisation" appear on certificates that use it. It is not a defect; it is a method with a stated assumption, which is that everything present absorbs equally at the detection wavelength. The assumption is false for peptides and the arithmetic proceeds regardless.

R. Steensen, Vejle

The Journal replies

Named assumptions are the whole argument of this department. A method that states what it is assuming can be argued with, and a method that leaves the assumption implicit cannot.

Our practice is to require the chromatogram alongside the number, and the striking thing is how often the request is met and how rarely anyone makes it. Two years of asking and we have been refused twice. The data is sitting on a laboratory server that nobody is querying.

E. Vasquez-Rueda, Cali

Your worked example varies gradient and threshold together and reports a 1.8-point spread. Which of the two contributed more? The article does not say, and the answer matters for what you are asking suppliers to disclose first.

M. Ferrari, Trieste

The Journal replies

Gradient, by roughly two to one in our four conditions: holding the threshold at 0.10 per cent, lengthening the gradient cost 0.9 points, while holding the gradient and tightening the threshold cost 0.4 to 0.9 depending on which gradient. We should have printed that decomposition in the table and it now appears in the note. If a supplier will disclose only one value, it should be the gradient.

Your submission design has a hole in it. Eight vials from one lot cannot separate variation between laboratories from variation between vials, because you have no replicate within a laboratory to estimate the second. Two vials each is a start and it is not enough, and the honest conclusion from your table is that the four figures differ, not that the laboratories do.

M. Bogdanović, Podgorica

The Journal replies

Correct, and the criticism is well aimed. With pairs we can see within-laboratory agreement, which was good in every case, but we cannot decompose the remaining variance properly. The four-condition study on a single sample was designed to isolate the method effect for exactly that reason, and it is the stronger half of the exercise. We should have said which half carried the weight.

Accreditation is worth less than buyers think and more than your article allows. A scope certificate tells you the laboratory has been audited for the methods listed on it, and nothing about whether your sample was run under one of them. Ask for the scope, then ask whether your test sits inside it.

J. Kiptoo, Eldoret

The Journal replies

That is the right two-step and we have adopted the phrasing. A laboratory can be accredited and still run your sample under a method outside its scope, entirely legitimately, and the report will not say so unless you ask.